Analysis of the Natural Organic Compound SAAF by Using the TOF-TOF Option

Introduction

Sperm activating and attracting factor (SAAF) is a naturally occurring substance produced by ascidians (sea squirts) to promote the fertilization process. Previously, the structure of this compound was characterized by high-energy collision induced dissociation (CID) using a traditional four sector tandem double focusing mass spectrometer.1

In this work, we analyzed the SAAF from a ciona intestinalis by using a JMS-S3000 SpiralTOF™ equipped with the optional TOF-TOF analyzer to generate a high energy CID product ion spectrum for this compound.

Results and Discussion

Fig. 1 shows the SAAF mass spectrum acquired using PEG sulfate 600 as an external calibrant. The spectrum shows a peak at m/z 515.301, which is suspected to be the monoisotopic ion [M-2Na-SO3+H]- (calculated value 515.305). In addition, peaks from SAAF are observed at two other masses: [M-2Na+H]- at m/z 595.255 (calculated value 595.261) and [M-Na]- at m/z 617.237 (calculated value 617.242). Among these peaks, the product ion spectrum was acquired for the [M-2Na-SO3+H]- ion, which had been previously analyzed using a traditional four sector tandem double focusing mass spectrometer.1 Fig. 2 shows the high energy CID results generated by the SpiralTOF-TOF. The negative charge was fixed to the sulfate ester end of the molecule, causing charge remote fragmentation (CRF) to occur. As a result, the spectrum showed peaks that systematically reflected the structure of SAAF, as shown in Fig. 3. This data closely resembles the results obtained using a traditional four sector MS/MS system.1

Attached Files
Showing 0 Comment


Comments are closed.

Other Resources

Walkup NMR
  • See how the Delta NMR software allows users to just "walk up" and start NMR experiments
  • Mass Spec Reference Data
  • View our page of useful molecular references for Mass Spec
  • Tutorials (Mass Spec)
  • Documents on the basics of mass spectrometry
  • Delta NMR software Tutorials
  • Videos on how to use the Delta NMR software
  • No-D NMR
  • Description of No-D NMR and how it can be used to eliminate the need for deuterated solvents
  • Non Uniform Sampling (NUS)
  • Description of how NUS is used to greatly reduce the time needed for running NMR experiments
  • NMR Basics
  • Overview of the Basics of NMR Theory
  • NMR Magnet Destruction
  • See our presentation of the slicing open of a JEOL Delta-GSX 270 MHz NMR Magnet
  • NMR Training
    Basic Operations and System Management for JEOL NMR Users
    Mass Spec Training
    Learn more about spectrometer operation and maintenance, data collection and processing, and advanced MS software operation.
    JEOLink NMR Newsletter
    We publish and send out this NMR newsletter to our customers. They can also be viewed here.
    Mass Media Newsletter
    We publish and send out this Mass Spec newsletter to our customers. They can also be viewed here.
    © Copyright 2024 by JEOL USA, Inc.
    Terms of Use
    |
    Privacy Policy
    |
    Cookie Preferences