Evolved gas monitoring from PA66 by using PY-GC-HRTOFMS

Introduction

Compounds which are generated upon heating polymer materials include monomers, additives and fragments characteristic of the polymer materials. Low molecular weight compounds generated by pyrolysis and/or locally desorption are observed. The relationship between the behavior of evolved gas compounds and the heating temperature is important for understanding the chemical characteristics of polymer materials. A gas chromatograph (GC) - mass spectrometer (MS) with thermogravimetric / differential thermal analyzer (TG/DTA) or pyrolyzer (Py) is generally used to identify the thermally evolved gas compounds from polymer materials. However, GC with a low-resolution MS such as quadrupole MS (QMS) may have difficulty monitoring low molecular-weight compounds generated by heating, since the GC is not used to provide chromatographic separation and QMS has poor mass spectrometric separation in the low mass region. (For example, N2+·, CO+· and C2H4+· are detected as the same signal of nominal mass m/z 28.)

We report the use of Py - GC - high resolution time-of-flight MS (HRTOFMS) for monitoring the low-molecular weight compounds generated upon heating the polymer materials. This report focuses on the generation of low molecular weight compounds near m/z 18, 28, and 4.

Experimental

A small piece of commercial PA66 (Nylon 66) 0.28mg was measured without further sample preparation by using PY/GC/HRTOFMS, using the JMS-T200GC "AccuTOF GCx-plus” coupled with a PY-2020iD. Measurement conditions were shown in Table 1. To detect evolved gases by heating, we used a low filament current.

Attached Files
Showing 0 Comment


Comments are closed.

Other Resources

Walkup NMR
  • See how the Delta NMR software allows users to just "walk up" and start NMR experiments
  • Mass Spec Reference Data
  • View our page of useful molecular references for Mass Spec
  • Tutorials (Mass Spec)
  • Documents on the basics of mass spectrometry
  • Delta NMR software Tutorials
  • Videos on how to use the Delta NMR software
  • No-D NMR
  • Description of No-D NMR and how it can be used to eliminate the need for deuterated solvents
  • Non Uniform Sampling (NUS)
  • Description of how NUS is used to greatly reduce the time needed for running NMR experiments
  • NMR Basics
  • Overview of the Basics of NMR Theory
  • NMR Magnet Destruction
  • See our presentation of the slicing open of a JEOL Delta-GSX 270 MHz NMR Magnet
  • NMR Training
    Basic Operations and System Management for JEOL NMR Users
    Mass Spec Training
    Learn more about spectrometer operation and maintenance, data collection and processing, and advanced MS software operation.
    JEOLink NMR Newsletter
    We publish and send out this NMR newsletter to our customers. They can also be viewed here.
    Mass Media Newsletter
    We publish and send out this Mass Spec newsletter to our customers. They can also be viewed here.
    © Copyright 2024 by JEOL USA, Inc.
    Terms of Use
    |
    Privacy Policy
    |
    Cookie Preferences