Integrated Analysis of Coffee Aroma by using a Headspace GC-HRMS - MSTips 280

Developing an Integrated Analysis Technique using Data Acquired by GC/EI and GC/Soft Ionization

Product: JMS-T200GC GC /MS System

Introduction

Electron ionization (EI) is a hard ionization method that is commonly used with gas chromatography mass spectrometry (GC-MS). The mass spectral fragmentation patterns produced by EI are used for library database searches to identify compounds. Conversely, soft ionization methods like field ionization (FI) tend to produce clear molecular ions with minimal fragmentation. When high-resolution MS is used with these ionization techniques, the accurate masses for the fragment ions produced by EI and the molecular ions produced by soft ionization provide an additional dimension of information for the analytes. Combining the exact mass information with the results of conventional library search can enhance the accuracy of identification compared to the use of library search alone.  In this work, we introduce the msFineAnalysis software and use it to automatically combine data acquired by GC/EI and GC/soft ionization for the qualitative analysis of coffee headspace.

Experiment

A commercial coffee was prepared as follows:

  1. One gram of coffee beans was loaded into a 22 mL HS vial, 15 mL of 100˚C water was added, and the vial was sealed.
  2. After the sample was cooled to room temperature, 10 mL of the supernatant was loaded into a HS vial, and 2 µL of an internal reference (p-Bromofluorobenzene) was added to the sample.
  3. Finally, 2 mL of the above solution was transferred and sealed in a vial that was then used as a sample.

Table 1 shows the measurement conditions used for the headspace/GC-TOFMS system.

Table 1. Measurement conditions

[Headspace Conditions]
System MS-62070STRAP (JEOL)
Mode Trap mode
Extract 3 times
Heating condition 60°C, 15 min
[GC-TOFMS Conditions]
System JMS-T200GC (JEOL)
Ionization mode EI+: 70 eV, 300 μA
FI+: -10 kV, 8mA (Carbotec 5 mm)
GC column ZB-WAX, 30 m x 0.18 mm, 0.18 μm
Oven temp. 40°C (3 min) → 30°C/min → 250°C (10 min)
Inlet temperature 250°C
Inlet mode Split 30:1

Results and discussions

Figure 1 shows the operational flow chart for the integrated analysis steps used for the JEOL msFineAnalysis software (chart on the right). First, the data is acquired by using both EI and soft ionization (SI), and all peaks and associated mass spectra are detected in the chromatograms. Afterwards, the mass spectra produced by these ionization methods are linked using their retention times, and these linked mass spectra are recorded as single components. Next, the EI mass spectrum is used for the library database search (1), and the SI mass spectrum is used to identify the analyte molecular ion (2). Afterwards, the molecular ion is used for exact mass analysis to determine possible elemental compositions, and these candidate formulas are then filtered by using the EI library search results (3). Next, the molecular ion is subjected to isotopic pattern analysis to help further limit the candidate formulas (4). Each candidate formula is then used as a search constraint for the exact mass analysis of the EI fragment ions (5). If the molecular ion formula candidate is incorrect, the EI fragment ions will not result in many (if any) compositional formulas, thus indicating that the molecular ion formula is not a good candidate for that particular analyte. These results are then output as an integrated qualitative report (6).


Figure 1. Qualitative Analysis Flow


Figure 2. TIC chromatograms of coffee aroma acquired by a HS/GC/TOFMS


Figure 3. Integrated qualitative analysis results on msFineAnalysis

The msFineAnalysis Auto Analysis function detected 67 components in the GC/EI and GC/FI measurements (Figure 2) that were automatically linked using their retention time. The Auto Analysis function then automatically used the steps in Figure 1 to analyze the linked data, and the results were output as a color-coded table as shown in Figure 3. Each color indicates a level of confidence for the identity of the compound:

Green: A molecular formula candidate was uniquely identified.
Orange: Multiple molecular formula candidates were identified.
White: No significant molecular formula candidates were identified.

The components classified as orange or white can be further reviewed manually to potentially identify a unique candidate formula. In this example, the software was able to automatically determine a unique molecular formula for 63 of the 67 components in the coffee headspace sample. 

Conclusions

The msFineAnalysis software produces highly accurate qualitative analysis results by automatically combining the EI library search results and soft ionization (SI) molecular formula determinations. Additionally, this software makes it possible to determine molecular formulas for unknown components not registered in library (match factor score: low), which can not be identified by database search alone (Figure 1, left side). The effectiveness of the msFineAnalysis integrated analysis method effectiveness for GC/MS qualitative analysis was demonstrated by automatically determining molecular formulas from exact masses, regardless of the level of match factor score, to limit the candidate formulas.

Attached Files
Showing 0 Comment


Comments are closed.

Other Resources

Walkup NMR
  • See how the Delta NMR software allows users to just "walk up" and start NMR experiments
  • Mass Spec Reference Data
  • View our page of useful molecular references for Mass Spec
  • Tutorials (Mass Spec)
  • Documents on the basics of mass spectrometry
  • Delta NMR software Tutorials
  • Videos on how to use the Delta NMR software
  • No-D NMR
  • Description of No-D NMR and how it can be used to eliminate the need for deuterated solvents
  • Non Uniform Sampling (NUS)
  • Description of how NUS is used to greatly reduce the time needed for running NMR experiments
  • NMR Basics
  • Overview of the Basics of NMR Theory
  • NMR Magnet Destruction
  • See our presentation of the slicing open of a JEOL Delta-GSX 270 MHz NMR Magnet
  • NMR Training
    Basic Operations and System Management for JEOL NMR Users
    Mass Spec Training
    Learn more about spectrometer operation and maintenance, data collection and processing, and advanced MS software operation.
    JEOLink NMR Newsletter
    We publish and send out this NMR newsletter to our customers. They can also be viewed here.
    Mass Media Newsletter
    We publish and send out this Mass Spec newsletter to our customers. They can also be viewed here.
    © Copyright 2024 by JEOL USA, Inc.
    Terms of Use
    |
    Privacy Policy
    |
    Cookie Preferences