Quantitative Analysis of Pyrazole Pesticides in Tea Leaf by Using FastGC-HRTOFMS

Introduction

The FastGC method is a very useful technique for doing rapid GC analyses that result in extremely narrow chromatographic peaks over a shorter time period than traditional GC analyses. Additionally, time-of-flight mass spectrometers (TOFMS) are capable of very fast data acquisition in comparison with other types of mass spectrometers so they are well suited as the detector for the FastGC technique. Furthermore, when the TOFMS is capable of high resolution measurements, the resulting mass spectra contain accurate mass information that can be used to calculate the elemental compositions for each observed m/z.

In this application note, we describe the quanitative analysis of pyrazole pesticides (Fipronil, Ethiprole, Pyraflufen ethyl and Tebfenpyrad) on tea leaves by FastGC/HRTOFMS. Additionally, we confirm that a rapid analysis with high sensitivity is easy to perform and very useful for fast screening.

Method

The tea leaf sample (5g) was prepared using the multiresidue method for agricultural chemicals by GC/MS published by Ministry of Health, Labour and Welfare, Japan. The pyrazole pesticides were added to make 0.01, 0.05 and 0.1ppm solution in the prepared solution from tea leaf. These concentrations in solution are equivalent to 4, 20 and 40ppb in tea leaf. Each sample was analyzed 3 times to check the reproducibility of the results.

Attached Files
Showing 0 Comment


Comments are closed.

Other Resources

Walkup NMR
  • See how the Delta NMR software allows users to just "walk up" and start NMR experiments
  • Mass Spec Reference Data
  • View our page of useful molecular references for Mass Spec
  • Tutorials (Mass Spec)
  • Documents on the basics of mass spectrometry
  • Delta NMR software Tutorials
  • Videos on how to use the Delta NMR software
  • No-D NMR
  • Description of No-D NMR and how it can be used to eliminate the need for deuterated solvents
  • Non Uniform Sampling (NUS)
  • Description of how NUS is used to greatly reduce the time needed for running NMR experiments
  • NMR Basics
  • Overview of the Basics of NMR Theory
  • NMR Magnet Destruction
  • See our presentation of the slicing open of a JEOL Delta-GSX 270 MHz NMR Magnet
  • NMR Training
    Basic Operations and System Management for JEOL NMR Users
    Mass Spec Training
    Learn more about spectrometer operation and maintenance, data collection and processing, and advanced MS software operation.
    JEOLink NMR Newsletter
    We publish and send out this NMR newsletter to our customers. They can also be viewed here.
    Mass Media Newsletter
    We publish and send out this Mass Spec newsletter to our customers. They can also be viewed here.
    © Copyright 2024 by JEOL USA, Inc.
    Terms of Use
    |
    Privacy Policy
    |
    Cookie Preferences