Using Solid Phase Microextraction with AccuTOF-DART for Fragrance Analysis

Introduction

Solid phase microextraction (SPME) is a well established sampling technique that is often used to isolate volatile organic components in gaseous mixtures. Once the compounds have been collected, the SPME fibers are typically placed into a heated GC inlet which thermally desorbs these components into a GC-MS system for analysis. Normally, this analysis can take between 10 and 30 minutes to complete depending on the complexity of the samples. In this work, the Direct Analysis in Real Time (DART™) heated gas stream is used to desorb and directly introduce a SPME sample into a high-resolution mass spectrometer. This methodology produces comparable information to the traditional GC-MS technique but streamlines the results into only a few seconds of analysis time.

Experimental

A Supelco DVB/Carboxen/PDMS StableFlex SPME fiber was placed in an enclosed plastic bag with a banana for 10 minutes during each analysis. For direct analysis of the SPME fiber, the JEOL AccuTOF-DART™ system was set to the following parameters: needle voltage 3500V, discharge electrode 150V, grid electrode 250V, helium temperature 200 degrees C, and helium flowrate 2.3 L/min. A JEOL GC-Mate II high resolution sector bench top system equipped with a DB5-HT (0.25mm × 30m) was used for the GC-MS portion of the analysis. The GC-Mate II was set to the following parameters: inlet temperature 250 degrees C, split ratio 30, and helium flowrate 1.2 mL/min. The GC oven was set for the following temperature profile: 40 degrees C held for 2 min, ramp from 40 to 260 degrees C at 20 degrees C/min, 260 degrees C held for 2 min.

Attached Files
Showing 0 Comment


Comments are closed.

Other Resources

Walkup NMR
  • See how the Delta NMR software allows users to just "walk up" and start NMR experiments
  • Mass Spec Reference Data
  • View our page of useful molecular references for Mass Spec
  • Tutorials (Mass Spec)
  • Documents on the basics of mass spectrometry
  • Delta NMR software Tutorials
  • Videos on how to use the Delta NMR software
  • No-D NMR
  • Description of No-D NMR and how it can be used to eliminate the need for deuterated solvents
  • Non Uniform Sampling (NUS)
  • Description of how NUS is used to greatly reduce the time needed for running NMR experiments
  • NMR Basics
  • Overview of the Basics of NMR Theory
  • NMR Magnet Destruction
  • See our presentation of the slicing open of a JEOL Delta-GSX 270 MHz NMR Magnet
  • NMR Training
    Basic Operations and System Management for JEOL NMR Users
    Mass Spec Training
    Learn more about spectrometer operation and maintenance, data collection and processing, and advanced MS software operation.
    JEOLink NMR Newsletter
    We publish and send out this NMR newsletter to our customers. They can also be viewed here.
    Mass Media Newsletter
    We publish and send out this Mass Spec newsletter to our customers. They can also be viewed here.
    © Copyright 2024 by JEOL USA, Inc.
    Terms of Use
    |
    Privacy Policy
    |
    Cookie Preferences